Refer to the urinalysis results below:
Color Yellow
Clarity Clear
Specific Gravity 1.010
pH 7.0
Protein Neg
Glucose Trace
Ketones Neg
Blood Neg
Bilirubin Neg
Urobilinogen 0.2 mg/dL
Nitrate Neg
Leukocytes Neg
Microscopic
Examination
WBC 25-50
RBC >100
Bacteria 3+
What typically would cause the discrepancy
between the reagent strip and microscopic
results?
a. Instrument out of calibration
b. pH is too alkaline
c. Microscopic accidently performed on different specimen
d. Specimen dilution



Answer :

Answer:The correct answer is d. Specimen dilution.

The discrepancy between the reagent strip and microscopic results can be attributed to specimen dilution. Here's why:

- The reagent strip shows a specific gravity of 1.010, which indicates a diluted specimen.

- The microscopic examination shows a high number of red blood cells (>100) and white blood cells (25-50), which is inconsistent with the reagent strip results.

- The reagent strip shows negative or trace amounts of protein, glucose, ketones, and blood, which is inconsistent with the microscopic findings.

Specimen dilution can cause:

- False-negative results on the reagent strip

- Incorrect specific gravity readings

- Inconsistent results between the reagent strip and microscopic examination

The other options are less likely to cause the discrepancy:

a. Instrument out of calibration: This would affect all results, not just the discrepancy between reagent strip and microscopic results.

b. pH is too alkaline: The pH is within the normal range (7.0), and an alkaline pH would not typically cause this discrepancy.

c. Microscopic accidentally performed on a different specimen: This is unlikely, as the results are from the same urinalysis test.

Note: This question requires knowledge of urinalysis and laboratory testing. If you have any further questions or need clarification, feel free to ask.